- Key Details
- Excellent accuracy of less than 1% reading of flow
- Atmospheric pressure, test circuit pressure and temperature measurement with the option to correct to standard conditions
- Cost effective, economical, entry level flow instrument with very high accuracy
- Precision pressure regulator control
- Air and gas flow ranges from 0.01 ml/min to 10,000 l/min. Pressures up to 14 bar and vacuum
Many Applications
Flow testing cookers, boilers, fires in tradition gases and hydrogen. The international standard tolerance on these appliances is +/- 5%, with an accuracy of 1% the FCO732 is prefect for the task.
Critical testing of respiration circuits, also known as a breathing circuit, a system that directs the flow of breathing gas to or from a user. Included in these are breathing tubes, breathing filter, anaesthetic machine and more. Flow measurement is essential for delivery of medical gases. Also, for checking for occlusions, by measuring the passage of air through an object, signifying if the object's passage is clear or occluded/blocked.
Incorporating temperature compensation and optional barometric pressure compensation to an already highly accurate flow measuring system.
Precision in Flow Testing
The FCO732 is a highly accurate flow meter for air and a wide range of other gases. It is used in many different production and laboratory applications. Calibrated to ISO 17025 standards to less than 1% accuracy of reading using a laminar flow element from flows of 0.01 ml/min to 10000 l/min. Automatic pressure and temperature compensation add still further to improve the accuracy.